Development of high-sensitivity radon detector in water for neutrino physics

Atsushi Takeda*, Yuuki Nakano, Kohei Okamoto, Yasuo Takeuchi

*この論文の責任著者

研究成果: ジャーナルへの寄稿会議記事査読

抄録

The contamination of radon (222Rn) originated from the decay chain of the uranium series in water could be serious background source in underground particle physics experiments. We have been developing a system for continuous monitoring of radon in water. We developed a new high-sensitivity radon detector for water by combining an existing high-sensitivity 80 L radon detector for air and membrane degasifier module used for previous version of radon detector for water. The housing component for membrane degasifier has been improved. The background level of the new radon detector for water was measured and it was reduced by approximately half compared to the previous one.

本文言語英語
論文番号012243
ジャーナルJournal of Physics: Conference Series
2156
1
DOI
出版ステータス出版済み - 2022/02/21
イベント17th International Conference on Topics in Astroparticle and Underground Physics, TAUP 2021 - Virtual, Online
継続期間: 2021/08/262021/09/03

ASJC Scopus 主題領域

  • 物理学および天文学一般

フィンガープリント

「Development of high-sensitivity radon detector in water for neutrino physics」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。